6. Commissioning, operation

6.3.2 Filling the micro calibration bath

1. Firstly, remove the transport cover.

2. Place the test item in the sensor basket.

3. Fill the tank with calibration liquid.

Following maximum filling heights are recommended:

Calibrator model

max. filling height

CTB9100-165 / CTM9100-150

130 mm

CTB9100-165 / CTM9100-150

with removable insert

110 mm

CTB9100-225

123 mm

CTB9100-225

with removable insert

105 mm

CTB9100-225-X

115 mm

CTB9100-225-X

with removable insert

95 mm

Following points should be observed at the maximum

filling heights:

Measured from the bottom of the sensor basekt

Non loaded tank

WIKA standard filling liquid

Pre-filling ex-works with the optimal height

The transport cover is fitted with a safety

valve. If the micro calibration bath is closed

when it is still warm, excessive pressure

can build up. In order to avoid overpres-

suring, which could lead to damage to the

liquid bath, the safety valve is activated at a

pressure of approx. 2.5 bar. This can lead to

hot vapour escaping.

6.3.3 Operation of the magnetic stirrer

The greatest possible homogeneity is achieved by the agita-

tion of the calibration liquid via a magnetic stirrer.

Set the stirring speed to the maximum possible. Turning the

wheel upwards will increase the speed, turning it downwards

slows down the stirring speed.

The magnetic stirrer is a

consumable.

Liquid bath

Front of the controller with stirrer-speed dial

6.3.4 Insert for liquids

The insert for liquids consists of:

Insert with leak-proof cover

Sensor basket

Magnetic stirrer and lifter

Replacement tools

If the insert for liquids is ordered with a new

model CTB9100 micro calibration bath or a

new model CTM9100 multi-function calibra-

tor, the instrument will have been matched to

the insert for liquids.

Should the insert for liquids be ordered after-

wards, a re-adjustment of the instrument at

the customer’s will be needed.

WARNING!

The insert for liquids must only be removed if

the calibrator is used at room temperature.

CAUTION!

The fill level must be matched the relevant

medium and the temperature.

Insert for liquids