13. Heating and cooling times

13.7 Micro calibration bath, model CTB9100-225
Measuring conditions

All times refer to a Pt100 reference sensor with Ø 6 mm

The reference sensor is positioned centrally in the tank,

5 mm over the strainer insert

All times are transition times and do not account for the

required settling time

The measurements were carried out at a room tempera-

ture of approx. 23 °C, with the cover off the bath

Distilled water

Heating up:

Time

Cooling:

Time

25 °C to 40 °C

0:55 min

90 °C to 75 °C

5:53 min

40 °C to 50 °C

0:37 min

75 °C to 50 °C

15:17 min

50 °C to 75 °C

1:27 min

50 °C to 40 °C

10:50 min

75 °C to 90 °C

1:30 min

40 °C to 25 °C

45:26 min

Silicone oil 5 CS

Heating up:

Time

Cooling:

Time

25 °C to 40 °C

0:51 min

120 °C to 100 °C

3:27 min

40 °C to 50 °C

0:16 min

100 °C to 75 °C

5:55 min

50 °C to 75 °C

0:54 min

75 °C to 50 °C

10:00 min

75 °C to 100 °C

1:13 min

50 °C to 40 °C

7:02 min

100 °C to 120 °C

1:35 min

40 °C to 25 °C

34:28 min

Silicone oil 10 CS

Heating up:

Time

Cooling:

Time

25 °C to 40 °C

0:52 min

165 °C to 150 °C

1:40 min

40 °C to 50 °C

0:22 min

150 °C to 125 °C

3:17 min

50 °C to 75 °C

0:52 min

125 °C to 100 °C

4:14 min

75 °C to 100 °C

0:53 min

100 °C to 75 °C

5:59 min

100 °C to 125 °C

0:59 min

75 °C to 50 °C

9:59 min

125 °C to 150 °C

1:12 min

50 °C to 40 °C

7:00 min

150 °C to 165 °C

1:03 min

40 °C to 25 °C

31:40 min