Technical manual – Merit Medical Worley Advanced CSG Coronary Sinus Guide Benutzerhandbuch

Seite 3

Advertising
background image

Technical Manual

en Hemostatic Tear-away Introducer System

with Infusion Side Port

da Haemostatisk skræl-væk indførings system med

infusions sideport

de Hämostatisches Abreiβ-Einführsystem mit

Infusionseitenport

el Σύστημα Αιμοστατικής Αποκόλλησης Εισαγωγέα με

‘Έγχυση Πλευρικής Θύρας

es Sistema introductor hemostático desechable con

puerto lateral para infusión

fi Infuusiosivuportilla varustettu hemostaattinen

irtirevittävä sisäänvientijärjestelmä

fr Système d’introduction hémostatique à détacher

avec orifice latéral de perfusion

it Sistema introduttore asportabile emostatico con

porta di infusione laterale

ja

引き剥がし型止血用導入器装置

(注入サイドポート付き)

nl Hemostatisch verwijderbaar inbrengsysteem met

zijopening voor infusie

no Hemostatisk avtakbar introduksjonssystem med

infusjonssideport

pt Sistema introdutor hemostático com orifício lateral

de infusão, que se retira rasgando

sv Hemostatiskt avdragbart introducersystem med

infusionssidoport

en Directions for Use

3

da Brugsvejledning

4

de Gebrauchsanweisung

5

el

Οδηγίες χρήσης

6

es

Instrucciones de uso

7

fi

Käyttöohje

8

fr

Mode d’emploi

9

it

Istruzioni per l’uso

10

ja

使用上の注意

12

nl

Gebruiksaanwijzing

13

no Bruksanvisning

14

pt

Instruções de Utilização

15

sv

Bruksanvisning

16

For U.S.-California Only.

Proposition 65, a State of California voter initiative,

requires the following notice:

WARNING: This product and its packaging have been ster-

ilized with ethylene oxide. This packaging may expose you to

ethylene oxide, a chemical known to the state of California

to cause cancer or birth defects or other reproductive harm.
CAUTION: Federal (U.S.A.) Law restricts this device to sale by

or on the order of a physician.
The hemostatic splittable sheath is protected by Patent In-

surance. USA Patent 7,462,184. European Patent EP1360972.

Other USA and Worldwide Patents pending.

en-Directions for Use

This device is intended for one time use only. Read

instructions prior to use.
Indications

For the introduction of various types of pacing or

defibrillator leads and catheters.
Contraindications

Use of the Coronary Sinus Guide / Lateral Vein Introducer

(CSG/LVI) systems is contraindicated for the following:

• Patients with an existing or possible occlusion of the

coronary vessels or unsuitable anatomy of the coronary

veins

• Patients with active systemic infection
Possible Negative Side Effects / Adverse Events

Coronary Sinus Guide / Lateral Vein Introducer (CSG. LVI)

systems should be used by physicians familiar with

percutaneous catheter introduction. Complications which

may be associated with the use of catheter introducer

systems include, but are not limited to, the following:

• Air embolism

• Allergic reaction to contrast media

• Arterial wall damage

• Bleeding

• Cardiac arrhythmias

• Cardiac tamponade

• Chronic nerve damage

• Damage to the heart valves

• Hematoma at the puncture site

• Infection

• Local tissue response, fibrotic tissue formation

• Myocardial damage

• Myocardial infarction

• Plaque dislodgement

• Pneumothorax

• Stroke and death

• Thrombus formation/emboli

• Vascular occlusion

• Vascular spasm

• Venous or cardiac perforation
Warnings

• This product is sensitive to light. Do not use if stored

outside the protective outer carton. Store in a cool, dark,

and dry place.

• Infusion through the side port can be done only after all

air is removed from the unit. Improper use of the

transvalvular insertion tool (TVI) can cause air embolism

and back bleeding.

• Do not use this device in patients who cannot be

appropriately anticoagulated. When tested in non-

anticoagulated sheep, this device has shown thrombus

formation, however, heparinized studies alleviated

the concern.
Precautions

• Do not alter this device in any way.

• Single Use Devices: This single-use product is not

designed or validated to be reused. Reuse may cause a

risk of cross-contamination, affect the measurement

accuracy, system performance, or cause a malfunction

as a result of the product being physically damaged

due to cleaning, disinfection, re-sterilization, or reuse.

• Federal law (U.S.A.) restricts this device to sale by or

on the order of a physician.

• Aspiration and saline flushing of the sheath, dilator,

and valve should be performed to help minimize the

potential for air embolism and clot formation.

• Indwelling introducer sheaths should be internally

supported by a catheter, pacing lead, or dilator.

• Dilators, catheters, and pacing leads should be removed

slowly from the sheath. Rapid removal may damage the

valve members resulting in blood flow though the valve.

Never advance or withdraw guide wire or sheath when

resistance is met. Determine cause by fluoroscopy and

take remedial action.

• When injecting or aspirating through the sheath, use

the side port only.

• When using the transvalvular insertion tool (TVI),

lead size may not exceed 6.2F.

• When using the TVI always keep the exposed

proximal end covered to prevent air embolization

and back bleeding.
Use Sterile Technique A suggested procedure:
1. Peel open package and place contents on sterile field.
2. Prep skin and drape in area of anticipated venipuncture

as desired.
3. Distend the subclavian vein. The subclavian vein is difficult

to locate unless it is distended by raising the patient’s legs

to a 45 degree angle or by using the Trendelenburg position.

The vein will be much easier to locate if the patient is well

hydrated.
4. Insert needle into vessel. The needle position should

be verified by observing venous blood return.
5. The angle of the needle should be adjusted depending

on the patient’s build: shallow in a thin person, deeper in a

heavy-set person. Use an 18g needle, 7cm (2-3/4 in.) long.
6. Aspirate the puncture needle using the 12cc syringe.
7. Remove the syringe and insert soft tip of guide wire

through the introducer needle into the vessel. Advance

guide wire guide to required depth. Leave an appropriate

amount of guide wire exposed. At no time should the guide

wire be advanced or withdrawn when resistance is met.

Determine the cause of resistance before proceeding. Fluoro-

scopic verification of the guide wire’s entrance into the

superior vena cava and right atrium is suggested.
8. Hold guide wire in place and remove introducer

needle. Do not withdraw the guide wire back into

the cannula as this may result in separation of the

guide wire. The cannula should be removed first.
9. Insert the straight vessel dilator into the sheath until

the dilator cap folds over the valve housing and

secures the dilator onto sheath assembly.
10. Thread the dilator/sheath assembly over the guide wire.
11. Advance the dilator and sheath together with a twisting

motion over the guide wire and into the vessel. Fluoroscopic

observation is advisable. Attaching a clamp or hemostat to

the proximal end of the guide wire will prevent inadvertent-

ly advancing the guide wire entirely into the patient.
12. Once assembly is fully introduced into the venous

system, separate the dilator cap from the sheath valve

housing by rocking the dilator cap off the hub.

(see Figure 1)
13. Slowly retract the dilator, leaving the sheath and wire in

position. The hemostasis valve will reduce the loss of blood

and the inadvertent aspiration of air through the sheath.
14. Remove the curved braided core from the package and

thread the exposed proximal portion of the retained

guide wire into the distal end of the braided core.
15. Feed the guide wire through the curved braided core or

braided sheath until the proximal end of the guide wire can

be secured with either a clamp or hemostat before advanc-

ing the curved dilator into the indwelling sheath.
16. Do not advance the braided core into the sheath until

the guide wire has been completely passed through the core

and the wire is secured with a hemostat or clamp in order

to prevent inadvertently advancing the guide wire entirely

into the patient.
17. Advance the braided core into the sheath and observe

fluoroscopically as the wire and distal end of the core extend

past the distal end of the sheath and are positioned in the

right atrium.
18. Manipulate the distal end of the guide wire or sheath

into the desired location (coronary sinus etc.) by combining

a twisting motion of the guide wire or sheath with the

gentle probing of the guide wire or sheath itself. Fluoroscopy

in the left anterior oblique (LAO) position is helpful. Advance

the CSG sheath into the mid coronary sinus and establish its

position by injecting contrast material through the side port.
19. Once the guide wire is in the desired location advance

the sheath over the wire until the tip rests in the desired

location. It is advisable to leave a short segment of wire

extending past the distal end of the tip to minimize any

3

Advertising