Checking the volume, Volumen kontrollieren – BrandTech Scientific Transferpette -8/-12 Benutzerhandbuch
Seite 15

6. Den zuvor gemessenen Istwert
einstellen. Volumeneinstellknopf
ca. 2 mm (ohne zu drehen) bis zum
Anschlag einschieben.
7. Distanzscheibe wieder über die
Achse schieben.
8. Abwerferhaube (A) montieren.
9. Abwerferhaube durch Drehen der
seitlichen Verschlüsse in Position
sichern.
6. Set to the previously measured
actual value. Push back volume
setting knob to the stop (approx.
2 mm), without turning it.
7. Replace the red distance plate
across the axis.
8. Mount ejector cap (A).
9. Secure ejector cap by turning the
lateral closures into position .
Checking the volume
The instrument should be calibrated at least once a
year. A more frequent interval may be dictated by
individual lab rules, applications and conditions.
The gravimetric volume test of the pipette requires
the following steps (page 16):
(Testing procedures are described e.g., in ISO 8655,
Part 6, “Testing Instruction”, page 37).
Volumen kontrollieren
Das Gerät sollte mindestens einmal jährlich überprüft
werden. Der Zyklus kann aber den individuellen An-
forderungen angepasst werden.
Die gravimetrische Volumenprüfung der Pipette erfolgt
durch folgende Schritte (Seite 16):
(Das Prüfverfahren ist z. B. in ISO 8655, Teil 6
beschrieben, „Prüfanweisung“, Seite 37).