Trouble shooting – BrandTech Scientific Transferpette -8/-12 Benutzerhandbuch
Seite 32

Problem
Possible Cause
Corrective Action
Pipette tip leaks or
volume too low
- Tip does not sit properly
- Press tip on firmly. Use only suitable
high-quality tips, page 8.
- Damaged V-ring (O-ring) on single
nose cone
- Replace V-ring (O-ring) on nose cones.
“Repairs”, pages 26-28.
- Piston damaged
- Replace piston.
“Repairs”, pages 24, 25.
- Seal dirty, damaged
- “Repairs”, pages 22, 23.
- Instrument is dirty
- Clean instrument.
“Cleaning”, page 19.
- Non-aqueous solutions pipetted
- “Adjustment”, page 14.
Poor sample aspiration
- Single nose cone clogged
- Clean nose cone. Remove cones only;
do not disassemble instrument!
“Cleaning” page 19.
Volume too large
- Pipetting key pressed too far down
into the blow-out stroke before
aspirating liquid.
- Operate pipette properly.
“Operating the pipette”, page 10.
Pipetting key moving
stiffly
- Key movement reduced
- Lubricate piston and centrale guide rod
with the appropriate grease,
pages 13, 34.
Trouble Shooting
- 32 -