Advertising
background image

3.2. Remote control commands

794 Basic Titrino

124

.Right:

Relative specification of the width of the output medium (e.g. pa-

per width) for the length of the measured value axis. 1 means the

measured value axis is plotted over the entire width of the paper

(largest possible width). In extreme cases, the writing of the right

tick may lie outside.

.Feed:

Length of the volume axis referred to the burette cylinder volume,

V(B) per cm (0.1 means, e.g. 1 mL/cm with a 10 mL Exchange

Unit). Depending on the printer, the measure in cm may not al-

ways be correct.

3.2.2.99. Setup.PowerOn $G

Simulation of 'power on'. The device has the same status as after power on:

The cylinder is filled, error messages deleted and the current sample number

set to 0. The method last used is ready for operation.

3.2.2.100. Setup.Initialise $G

Setup.Initialise.Select ActMeth, Silo, Calib, Config,

Assembly, Setup, All

Setting of default values for the following areas:

ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Calib:

pH calibration data for all measuring inputs.

Config:

All values under &Config.

Assembly: All values under &Assembly.

Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.

3.2.2.101. Setup.RamInit $G

Initializes instrument, see page 147. All parameters are set to their default

value and error messages are cleared. The user and silo memories will be de-

leted. The user memory contains the default user methods from Metrohm.

3.2.2.102. Setup.InstrNo $G

Setup.InstrNo.Value serial

number, 8 ASCII characters

Instrument identification for report output.

Set the value with &Setup.InstrNo $G .

3.2.2.103. Diagnose.Report $G

Output of the report containing the adjustment parameters. The Titrino has to

be in its inactive basic state.