Metrohm 794 Basic Titrino Benutzerhandbuch

Seite 87

Advertising
background image

3.1 General rules

794 Basic Titrino

83

RS send errors:

E40

DSR=OFF No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E41

DCD=ON No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E42

CTS=OFF No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E43

The transmission of the Titrino has been interrupted with XOFF for

at least 6 s.

Exit: Send XON or <QUIT>

E44

The RS interface parameters are no longer the same for both

devices. Reset.

E45

The receive buffer of the Titrino contains an incomplete command

(L

F

missing). Sending from the Titrino is therefore blocked.

Exit: Send L

F

or <QUIT>.


E120

Overrange of the primary measured value (pH, U, Ipol, Upol or T

with MEAS T). The secondary measured value (temperature) may

be instable as well.

Exit: Correct error or &m $S.

E121

Measuring point list overflow (more than 500 measuring points).

Exit: The error message disappears on next startup.

E122

EP overflow.

Exit: The error message disappears on next startup or on

recalculation.

E123

Missing EP for calculation.

Exit: The error message disappears on next startup or on

recalculation.

E124

Number of EP does not correspond with the set windows.

Exit: The error message disappears on next startup or on

recalculation.

E125

Missing fix EP for calculation, has not been defined.

Exit: The error message disappears on next startup or on

recalculation.

E126

Fix-EP outside of measuring point list.

Exit: The error message disappears on next startup or on

recalculation.

E128

No new mean.

Exit: The error message disappears on next startup or on

recalculation.

E129

No new common variable, old value remains.

Exit: The error message disappears on next startup or on

recalculation.

E130

Wrong sample. For SET with preset titration direction the first

measured value lies behind the endpoint.

Exit: The error message disappears on next startup.

E131

No EP set for SET.

Exit: The error message disappears on next startup.

E132

Silo empty and it has been started with open silo or empty silo has

been opened.

Exit: Send a silo entry.