3.1. General rules

794 Basic Titrino

84

E133

Silo full.

Exit: Send new command.

E134

No method. A method, which is required from the silo memory or in TIP, does

not exist.

Exit: The error message disappears on next startup.

E135

Check temp.sensor in MEAS T.

Exit: Correct error or &m $S.

E136

Same buffer in CAL. Measured value of the second buffer differs less than 6

mV from the measured value of the first buffer.

Exit: Correct error or &m $S.

E137

XXX Bytes are missing so that the method, the silo line could not be stored or

not enough RAM for running TIP.

Exit: Send new command.

E155

No new silo result (C24 or C25).

Exit: The error message disappears on next start or on recalculation.

E157

No sequence defined in TIP.

Exit. The error message disappears on next start.

E158

A second TIP has been called up in TIP.

Exit: The error message disappears on next start.

E160

No new temporary variable.

Exit: The error message disappears on next start.

E161

Measurement range of the secondary measured value (temperature)

exceeded. The primary measured value (pH, U, Ipol, Upol) can also be

unstable.

Exit: Rectify error or &m $S.

E166

Save lines is "OFF" although a submethod of TIP includes an assignment to

C24 or C25.

Exit: The error message disappears on next start. Attention: The data of this

sample will not be stored.

E172

In TIP, a QuickMeas was started, without defining a measuring quantity.

Exit: The error message disappears on next start or &Mode.QuickMeas $S.