Technical data

Temperature ranges:

75 ... 550°C (MB 5.5)

150 ... 1200°C (MB 12)

100 ... 700°C (MB 7)

200 ... 1800°C (MB 18)

Sub range:

any range adjustable within the temperature range
minimum span 51°C

Spectral range:

2 to 2.6 µm (main wavelength 2.3 µm)

IR detector:

Extended InGaAs

Power supply:

24 V DC (10 to 30 V DC), ripple must be less than 0.5 V

Power consumption:

Max. 1 W

Analog output:

0 to 20 mA or 4 to 20 mA (linear), switchable

Load:

0 to 500 Ω

Switch contact:

Opto relays; max. 50 V DC, 0.2 A; max. P

max

= 300 mW

Hysteresis:

2 to 20°C

Digital Interface:

RS485 addressable (half duplex), baud rate 1200 up to 38400 Bd

Resolution:

0.1°C on interface;
< 0.025% of the adjusted temperature sub range at the analog output

Isolation:

power supply, analog output and digital interface are galvanically isolated from
each other

Parameters:

Adjustable via interface:
Emissivity ε, transmittance

, exposure time t

90

, clear times for max. / min. value

storage or automatically deletion of max. / min. value storage, analog output
0 - 20 or 4 - 20 mA (switchable), sub temperature range, ambient temperature
compensation, address, switch contact, hysteresis, baud rate, wait time t

W

Emissivity ε:

10.0 to 100.0% adjustable via interface in steps of 0.1%

Transmittance

:

10.0 to 100.0% adjustable via interface in steps of 0.1%

Exposure time t

90

:

2 ms (with dynamical adaptation at low signal levels);

adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Maximum / minimum
value storage:

Built-in single or double storage. Clearing with adjusted time t

clear

(off; 0.01 s;

0.05 s; 0.25 s; 1 s; 5 s; 25 s), via interface or automatically with the next meas-
uring object

Uncertainty:

(with ε =1, t

90

=1 s, T

amb.

=23°C)

up to 400°C: 2°C
above 400°C: 0.3% of measured value in °C + 1°C
above 1500°C: 0.5% of measured value in °C

(Note: the pyrometer must be operate at least 30 min before these values are valid)

Repeatability:

(ε =1, t

90

=1 s, T

amb.

=23°C)

0.1% of measured value in °C + 1°C

Protection class:

IP65 (IEC 60529)

Ambient temperature:

0 to 70°C on housing, T

meas.

at least 25°C above T

housing

Storage temperature:

-20 to 70°C

Rel. humidity:

No condensing conditions

Weight:

0.3 kg

Housing:

Stainless steel

Connector:

8 pin connector

Dimensions [mm]:

Sighting:

Built-in LED targeting light

CE-label:

According to EU directives about electromagnetic immunity

Advertising