1 appropriate use, 2 scope of delivery, 2 safety – LumaSense Technologies IGA 320/23 Benutzerhandbuch

Seite 6: 1 general, 2 electrical connection, 3 electrical installation, Appropriate use, Scope of delivery, Safety, General

Advertising
background image

IMPAC pyrometer IGA 320/23

6

1.1

Appropriate use

The IGA 320/23 is a short wave infrared temperature measuring device with digital signal processing. It is
used for non-contact temperature measurements on metals, ceramics, graphite, etc. with 4 temperature
ranges between 75 and 1800°C.


1.2

Scope of delivery

Pyrometer IGA 320/23, PC adjustment and evaluation software “InfraWin”, inspection sheet, manual.

Note:

A connection cable is not included with the instrument and has to be ordered
separately (see section 12, Reference numbers).


2

Safety


This section offers an overview about important safety aspects. Additionally in the several sections there are
concrete safety aspects to avert danger. These aspects are indicated with symbols. Labels and markings at
the instrument have to be noticed and keep in a permanent readable condition.


2.1

General

Each person working with the pyrometer must have read and understood the user manual before operation.
Also this has to be done if the person already used similar instruments or was already trained by the manu-
facturer.
The pyrometer has only to be used for the purpose described in the manual. It is recommended to use only
accessories offered by the manufacturer.


2.2

Electrical connection

Follow common safety regulations for mains voltage (230 or 115 V AC) connecting additional devices operat-
ing with this mains voltage (e.g. transformers). Touching mains voltage can be mortal. A non expert connec-
tion and mounting can cause serious health or material damages.
Only qualified specialists are allowed to connect such devices to the mains voltage.

3

Electrical Installation


The IGA 320/23 is powered by a voltage of 24 V DC (possible range 10 to 30 V, ripple < 0.5 V). The instru-
ment needs no warm up time and is ready for use immediately.
For switching off the instrument, interrupt the power supply or unplug the electrical connector.

To meet the electromagnetic requirements (EMV), a shielded connecting cable must be used. The shield of
the connecting cable has to be connected only on the pyrometer’s side. On side of the power supply (switch
board) the shield must be open to avoid ground loops.

LumaSense offers connecting cables, they are not part of standard scope of delivery. The connecting cable
has wires for power supply, interface, analog output, switch contact and external clearing of the maximum
value storage via contact and 8 pin connector (see section 12, Reference numbers).