LumaSense Technologies IS 5/F Benutzerhandbuch

Seite 18

Advertising
background image

-

18

-

Optics
Target area size M [mm] when focussed to distance a [mm] with 90% of
radiation.

Range [°C]

600...1400

700...1800

800...2500

1000...3000

Aperture D [mm]

6.0

6.0

6.0

4.0

Focus to
measuring
distance

250 mm

6.0

3.0

1.5

1.5

300 mm

8.0

4.0

2.0

2.0

500 mm

11.0

5.5

2.8

2.8

800 mm

16.0

8.0

4.0

4.0

1300 mm

26.0

13.0

6.5

6.5

2000 mm

40.0

20.0

10.0

10.0

The target diameter changes with measuring distance, if the used distance
is different to the focal point distance a:

((

))

M

a

a

M

D

D

2

2

==

++

−−

((

))

M

a

a

M

D

D

1

1

==

−−

++

Ø M

Ø M

2

Ø M

1

a

a

1

Ø D

a

2

Advertising