6. Commissioning, operation

to select certain functions is limited and dependent upon what is already selected in the

other display sector. The "NEXT" option is used to change to the second parameter.

Only the resistance thermometer and pressure modes have a second parameter.

Resistance thermometers can be displayed in Celsius or Fahrenheit and for pressure

there are 11 different units.

Within any single display sector, the following modes are available:

P[1] = integrated pressure sensor

[EXT] = pressure with external pressure module

P[1] ST = switch test with integrated sensor

[EXT] ST = switch test with external pressure module

Current functions are only available on the lower display:

mA measurement = current measurement without voltage supply

mA w/24V = current measurement with voltage supply

mA Source = current source

mA sim = mA simulation using an external supply from the instrument under test

VOLT = voltage measurement

RTD = resistance thermometer (in the event a resistance thermometer is connected)

The following table shows which functions are available concurrently.

An "--" in a column indicates that the mode in the active display will not be available for

selection if the mode in that row is in use in any other display sector.

Active display

Other displa

ys

P[1] [EXT] P[1]

ST

[EXT]

ST

mA

mA-loop Volt

RTD

P[1]

[EXT]

P[1] ST

--

--

--

--

--

[EXT] ST

--

--

--

--

--

mA

--

--

--

--

mA-loop

--

--

--

--

Volt

--

--

--

--

RTD

6.1.2.3 Accessing other menus

In order to access further menus, select the "MORE" option in the main menu.