6. Commissioning, operation

To use the "%-ERROR" function, perform the following:

1. Once the calibrator is switched on and operating, press the F3 key to activate the

"MORE" menu option. Then press the F1 key to activate the "%-ERROR" option.

2. Press the F1 key to select the "CONFIG" option.

3. The first option is the port setting. Use the "SELECT" option to scroll through the

choice of ports (pressure connections). Once this is completed, select the "NEXT"

option.

4. "LOOP POWER" (voltage supply) can be toggled on or off; select "NEXT" when

done.

5. Use "SELECT" to scroll through the "UNIT" options, and select "NEXT" to move

on.