WIKA operating instructions, model CPH6600

14005103.03 11/2012 GB/D

GB

6. Commissioning, operation / 7. Maintenance, cleaning, ...

6.14 MIN/MAX memory

The calibrator has a MIN/MAX function for capturing the minimum and maximum values

of any displayed parameter.

The MIN/MAX function can be set by stepping through the menu options until "MIN/

MAX" is shown on the display above the F1 key. At this point, pressing the F1 key

will scroll the display through the MIN/MAX values that are stored in the "MIN/MAX"

registers. These readings are live so new MIN/MAX values will be recorded while in this

mode.

To clear the "MIN/MAX" registers, simply press the "CLEAR" key. These registers are

also cleared on power-up or when the configuration is changed.

7. Maintenance, cleaning, and recalibration

7.1 Maintenance

Repairs must only be carried out by the manufacturer.

Replacing the batteries

To avoid false measurements, replace the batteries as soon as the battery indicator

comes on. If the batteries run too low the CPH6600 will automatically shut down to avoid

the batteries leaking.

Only AA size alkaline batteries or rechargeable batteries, from the

optional CPH6600 accessories, should be used.

CAUTION!

To avoid personal injury or damage to the calibrator, only use

accessories supplied by WIKA and ensure that no water finds its way into

the case.