Faults display reason measures – WIKA CPH6600 Benutzerhandbuch

Seite 44

Advertising
background image

44

WIKA operating instructions, model CPH6600

14005103.03 11/2012 GB/D

GB

7. Maintenance, cleaning, and service / 8. Faults

7.3 Recalibration
DKD/DAkkS certificate - Certificates:

We recommend that the instrument is regularly recalibrated by the manufacturer, with

time intervals of approx. 12 months. Every factory recalibration includes, additionally,

an extensive free-of-charge check of all system parameters with respect to their

compliance with the specification. The basic settings will be corrected if necessary.

8. Faults

Display

Reason

Measures

Low battery power, device will

only work for a short while longer

Replace battery

OL

-OL

Well above or below the limits of

the measuring range

Check: is the pressure within the

permissible measuring range of

the sensor?

--> Reduce the applied pressure

No display or

strange symbols,

device is not

responding to

input

Battery flat

Replace battery

System error

Switch off the instrument, wait

for a short period of time, switch

on again

Technical fault

Return to manufacturer for repair

CAUTION!

If faults cannot be eliminated by means of the measures listed above,

the hand-held pressure calibrator must be shut down immediately, and it

must be ensured that pressure and/or signal are no longer present, and it

must be prevented from being inadvertently put back into service.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter

9.2 "Return".