Gb 6. commissioning, operation – WIKA CPH6600 Benutzerhandbuch
Seite 41
Advertising
41
WIKA operating instructions, model CPH6600
14005103.03 11/2012 GB/D
GB
6. Commissioning, operation
6. Use the arrow keys to set the upper limit of the measuring range; select
"DONE SET" when finished.
7. Use the arrow keys to set lower limit of the measuring range, and select
"DONE SET" when finished. The "%-ERROR" mode will be ready to use.
The lower and upper limit of the measuring range will be saved in
non-volatile memory until they are changed again by the user for the
internal pressure sensor and the external pressure modules.
Advertising