Gb 6. commissioning, operation – WIKA CPH6600 Benutzerhandbuch

Seite 39

Advertising
background image

39

WIKA operating instructions, model CPH6600

14005103.03 11/2012 GB/D

GB

6. Commissioning, operation

6.13.3 %-Error function

The calibrator features a special function which can calculate the error in the pressure

value from the mA value as a percentage of the 4 ... 20 mA span. The %-Error mode

uses all three screens and has a special menu structure. It displays pressure, mA and

%-Error simultaneously.

Example:

Suppose a pressure transmitter under test has a full scale range (measuring range) of

20 bar and gives a corresponding 4 ... 20 mA output signal. The user can programme in

a 0 ... 20 bar pressure span into the calibrator and the calibrator will then calculate and

display the deviation or %-Error value from the 4 ... 20 mA output. This then eliminates

manual calculations of the test item deviation.

Figure 6: CPH6600 with a pressure transmitter with %-Error function