Gb 3. specifications 3. specifications, Measuring ranges, Base instrument – WIKA CPH7600 Benutzerhandbuch
Seite 10

10
WIKA operating instructions, model CPH7600
14051628.01 11/2012 GB/D
GB
3. Specifications
3. Specifications
Measuring ranges
Pressure
Relative pressure
-0.8 ... +20 bar
Overpressure limit
40 bar
1)
Burst pressure
135 bar
1)
Accuracy
0.025 % of reading ±0.01 % FS
Resolution
5-digit
Temperature compensation
15 … 35 °C
Temperature coefficient
0.002 % of the span/°C beyond the 15 ... 35 °C
Current
Measuring range
0 … 24 mA (max. load 1,000 Ω)
Resolution
1 µA
Accuracy
0.015 % of reading ±2 µA (simulation and measurement)
Voltage
Measuring range
DC 0 … 30 V
Resolution
1 mV
Accuracy
0.015 % of reading ±2 mV (measurement)
1) Maximum pressure for the integrated lines: 40 bar
Base instrument
Pressure supply
maximum pressure for the integrated pressure generation: 20 bar
maximum pressure via external pressure supply: 7 bar
Pressure connection
1/8 NPT (female)
Permissible media
clean, dry, non-corrosive gases, compatible with silicon, Pyrex,
RTV, gold, ceramic, nickel and aluminium
Output
Voltage supply
DC 24 V
Power supply
Battery type
DC 16 V, NiMH rechargeable battery
Battery life (fully-charged)
approx. 50 hours (only measurement or with external pressure
supply)
125 pump cycles to 20 bar
300 pump cycles to 10 bar
1,000 pump cycles to 2 bar
Permissible ambient conditions
Operating temperature
-10 … +50 °C
Storage temperature
-20 … +60 °C
Relative humidity
35 ... 85 % r. h. (no condensation)