3. Specifications

3. Specifications

Measuring ranges

Pressure
Relative pressure

-0.8 ... +20 bar

Overpressure limit

40 bar

1)

Burst pressure

135 bar

1)

Accuracy

0.025 % of reading ±0.01 % FS

Resolution

5-digit

Temperature compensation

15 … 35 °C

Temperature coefficient

0.002 % of the span/°C beyond the 15 ... 35 °C

Current
Measuring range

0 … 24 mA (max. load 1,000 Ω)

Resolution

1 µA

Accuracy

0.015 % of reading ±2 µA (simulation and measurement)

Voltage
Measuring range

DC 0 … 30 V

Resolution

1 mV

Accuracy

0.015 % of reading ±2 mV (measurement)

1) Maximum pressure for the integrated lines: 40 bar

Base instrument

Pressure supply

maximum pressure for the integrated pressure generation: 20 bar

maximum pressure via external pressure supply: 7 bar

Pressure connection

1/8 NPT (female)

Permissible media

clean, dry, non-corrosive gases, compatible with silicon, Pyrex,

RTV, gold, ceramic, nickel and aluminium

Output
Voltage supply

DC 24 V

Power supply
Battery type

DC 16 V, NiMH rechargeable battery

Battery life (fully-charged)

approx. 50 hours (only measurement or with external pressure

supply)

125 pump cycles to 20 bar

300 pump cycles to 10 bar

1,000 pump cycles to 2 bar

Permissible ambient conditions
Operating temperature

-10 … +50 °C

Storage temperature

-20 … +60 °C

Relative humidity

35 ... 85 % r. h. (no condensation)

Advertising