Gb 6. commissioning, operation – WIKA CPH7600 Benutzerhandbuch

Seite 19

Advertising
background image

19

WIKA operating instructions, model CPH7600

14051628.01 11/2012 GB/D

GB

6. Commissioning, operation

Here the "SELECT" option will switch between the available choices for each

parameter. The first parameter is "MODE". Since voltage, current and switch test

modes all use the same leads, two of these functions cannot be used concurrently.

The ability to select certain functions is limited and dependent upon what is already

selected in the other display sector. The "NEXT" option is used to change to the second

parameter. Only the pressure modes have a second parameter.

Within any single display sector, the following modes are available:

P[1] = integrated pressure sensor

[EXT] = pressure with external pressure module

P[1] ST = switch test with integrated sensor

[EXT] ST = switch test with external pressure module

Current functions are only available on the lower display:

mA measurement = current measurement without voltage supply

mA w/24V = current measurement with voltage supply

mA Source = current source

mA sim = mA simulation using an external supply from the instrument under test

VOLT = voltage measurement

The following table shows which functions are available concurrently.

An "--" in a column indicates that the mode in the active display will not be available for

selection if the mode in that row is in use in any other display sector.

Active display

Other displa

ys

P[1] [EXT] P[1]

ST

[EXT]

ST

mA

mA-loop Volt

RTD

P[1]

[EXT]

P[1] ST

--

--

--

--

--

[EXT] ST

--

--

--

--

--

mA

--

--

--

--

mA-loop

--

--

--

--

Volt

--

--

--

--

6.1.2.3 Accessing other menus

In order to access further menus, select the "MORE" option in the main menu.