Gb 3. specifications, Available pressure range and resolution, Case – WIKA CPH7600 Benutzerhandbuch
Seite 11: Approvals and certificates, Pressure range and factors

11
WIKA operating instructions, model CPH7600
14051628.01 11/2012 GB/D
GB
3. Specifications
Case
Material
Case
NK-7TM Resin
Front panel
Aluminium
Ingress protection
IP 67 (case closed)
IP 40 (case opened)
Dimensions
387.4 x 304.8 x 177.8 mm (15.25 x 12 x 7")
Weight
approx. 7 kg (15.5 lbs.)
Approvals and certificates
CE conformity
EMC directive
2004/108/EC, EN 61326 emission (group 1, class B) and inter-
ference immunity (portable equipment)
Certificate
Calibration
3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
For further approvals and certificates, see website.
Available pressure range and resolution
Pressure range and factors
Relative pressure
-0.8 ... +20 bar
Overpressure limit
40 bar
Burst pressure
135 bar
Unit
Conversion factor
psi
1
300.00
bar
0.06894757
20.684
mbar
68.94757
20,684
kPa
6.894757
2,068.4
MPa
0.00689476
2.0684
kg/cm²
0.07030697
21.092
cmH
2
O (4 °C)
70.3089
21,093
cmH
2
O (20 °C)
70.4336
21,130
inH
2
O (4 °C)
27.68067
8,304.2
inH
2
O (20 °C)
27.72977
8,318.9
inH
2
O (60 °F)
27.70759
8,312.3
mmHg (0 °C)
51.71508
15,515
inHg (0 °C)
2.03602
610.81
For further specifications see WIKA data sheet CT 17.01 and the order documentation.