Gb 6. commissioning, operation – WIKA CPH7600 Benutzerhandbuch

Seite 20

Advertising
background image

20

WIKA operating instructions, model CPH7600

14051628.01 11/2012 GB/D

GB

6. Commissioning, operation

6.2 Using the backlighting

The backlight is controlled by the "LIGHT" option on the main menu. The light is toggled

on and off when the F3 key is pressed.

6.3 Using the "ZERO" function

When the ZERO key is pressed, the calibrator will zero the {active display}, as long as

a pressure mode is selected and the pressure is within the zero limit. The zero limits are

within 5 % of the full scale range of the selected sensor. If the display indicates "OL" the

zero function will not operate.

The ZERO key is only used for pressure.

6.3.1 External pressure module (not absolute)

When a pressure module is selected in the {active display} and the ZERO key is

pressed, the calibrator subtracts the current displayed reading from the output value.

The zero limits are within 10 % of the full scale range of the selected sensor. If the

display indicates "OL" the zero function will not operate.

6.3.2 External pressure module (absolute)

When an external absolute pressure module is selected in the {active display} and the

ZERO key is pressed, the calibrator prompts the user to enter the barometric reference

pressure. For this the arrow keys can be used. The sensor port should be opened

(vented) to atmosphere while performing this procedure.

6.4 Other menu controlled functions

There are 10 sub-menus that can be accessed through the "MORE" option of the main

menu. A sub-menu contains three options. The first option is unique to this function. The

second and third options of a sub-menu are always the same. The "NEXT" option leads

to the next sub-menu and the "DONE" option returns to home.

Advertising