6. Commissioning, operation

6.10 Performing a pressure switch test

Figure 3: CPH7600 with pressure switch

To perform a switch test, the following steps must be taken:

1. The upper display is set to [P1] ST, all other displays are switched off.

The pressure switch test can be performed with the following functions

[P1] ST or EXT ST.

2. Connect the calibrator to the pressure switch using the switch terminals. The polarity

of the terminals does not matter. Then connect the pressure side of the calibrator

and the pressure switch.

Test item

Output signal

DC 24 V voltage

COM

Pressure

switch

Use small-volume

hose lines, as for as

possible