Gb 6. commissioning, operation – WIKA CPH7600 Benutzerhandbuch

Seite 29

Advertising
background image

29

WIKA operating instructions, model CPH7600

14051628.01 11/2012 GB/D

GB

6. Commissioning, operation

6.7.2 Measuring pressure with external modules

The calibrator has a digital interface for external pressure modules. These modules

are available for various ranges, including gauge, vacuum, differential and absolute

pressure. The modules work seamlessly in conjunction with the calibrator. Simply

plug them into the interface and select [EXT] (external sensor). Since the interface

between the calibrator and the module is digital, the accuracy and display resolution is

dependent on the pressure module.

6.8 Measuring and sourcing current (4 ... 20 mA)

1. To measure current, use the input terminals on the front of the calibrator. Select the

mA function on the lower display. Current is measured in mA and percentage of the

measuring range. The measuring range on the calibrator is set to 0 % at 4 mA and

100 % at 20 mA.

Figure 2: Pressure measurement with external pressure module

Valve

Pressure module

Pressure module

adapter

Advertising