5 Configuration and parameter tables
Logic output 1 (open-collector output 1)
C221 Function of the logic output 1
X
0
no function
1
lk7
2
lk8
3
fault output
You will find additional information in
“Setting the limit values (limit comparators)” on page 16 and
C222 Signal of logic output 1 on
probe break/short-circuit
X
0
active
1
inactive
2
unchanged
Para-
meter
Explanation
Value range factory
setting
X
C223 Lower switching
differential (hysteresis)
0 — 250
100
The value range corresponds to 0 — 2.5%.
Para-
meter
Explanation
Value range factory
setting
X
C224 Upper switching
differential (hysteresis)
0 — 250
100
The value range corresponds to 0 — 2.5%.