Work procedures

Bitter units (EBC method)

Principle

Bitter substances, mainly iso-

 acids, are extracted with iso-octane from

the acidified sample and the concentration in the extract is determined
spectrophotometrically.

Application area

Beer, worts

Measurement range

10–40 bitter units (beer)
20–60 bitter units (wort)

Accessories

Centrifuge glasses with solvent-proof seals, 35 mL

Glass beads

Shaking machine

Centrifuge, 3000 rpm

Spectrophotometer, 275 nm

10 mm rectangular cuvette QS

Reagents

Hydrochloric acid, 6N

Iso-octane (2,2,4-trimethylpentane), spectroscopic pure (extinction
measured in 10 mm rectangular cuvette QS at 275 nm against dist.
H

2

O <0.010) (e.g. Uvasol)

Sample preparation

1.

Separate wort and cloudy beer through centrifuging at 3000 rpm
for 15 min (do not filter sample)

2.

Remove carbon dioxide from the beer without removing any foam.

3.

Pipette 10 mL (or 5 mL wort + 5 mL dist. H

2

O) of the sample

maintained at 20 °C (68 °F) into a centrifuge glass.

4.

Add 0.5 mL 6N hydrochloric acid, 20 mL iso-octane and
three glass beads.

5.

Seal the centrifuge glass and leave to shake mechanically for
15 min at 20 °C (68 °F).

6.

Centrifuge for 3 min at 3000 rpm.

7.

Measure the extinction of the iso-octane extract in 10 mm
rectangular cuvette at 275 nm against iso-octane of the same
quality (blank value).

Results specifications

Bitter units (BU) without decimals

Accuracy

r = -0.36 + 0.05 m
R = 0.72 + 0.14 m

Standard values

Beer: 10–40 BU, depending on nature, sort, type and origin
Wort: 20–60 BU, depending on beer and bitters yield

Literature

MEBAK brew-technical analysis methods
4th edition 2002, Volume II, P. 114 et seq