English 45

Determination of the calibration curve

The factor 1 = 0.037 is an empirical factor and must be individually
determined through a calibration curve. The factor is the slope of the
calibration curves.

1.

Pipette 40 mL of beer into four individual 50 mL volumetric flasks.

2.

Pipette into the volumetric flasks each0.40 mL, 0.80 mL, 1.60 mL
and 3.20 mL of the iron standard solution (5 mg Fe

3+

/mL).

3.

Add to each volumetric flask 2 mL ferrozine reagent and 1 mL
ascorbic acid solution.

4.

Fill to the mark with dist. H

2

O.

5.

Measure the extinction of the solution in a 40 mm rectangular
cuvette at 560 nm against a corresponding blank value.

Subtract the results of the un-spiked sample from the spiked samples.

Using these programs with the SIP 10 sipper

Information on the installation, configuration, and sample insertion when
using the SIP 10 sipper can be found in the SIP 10 user manual.

Accessories

Description

Order no.

Rectangular cuvettes set (standard deviation = 10 mm,
aligned pair)

2095100

Rectangular cuvette QS, standard deviation = 10 mm
(3.5 mL)

2624410

Rectangular cuvette OS, standard deviation = 10 mm
(3 units)

LZP045

Rectangular cuvette QS, standard deviation = 20 mm

LZV008

Rectangular cuvette OS, standard deviation = 20 mm

LZP331

Rectangular cuvette QS, standard deviation = 50 mm
(17.5 mL)

2624450

Rectangular cuvette OS, standard deviation = 50 mm

LZP167

Pour-through cuvette QS,

Standard deviation = 3 mm, center height = 10 mm, height =
40 mm

LZV638

Pour-through cuvette QS,

Standard deviation = 50 mm, center height = 10 mm, height
= 40 mm

LZV649

Pour-through cuvette QS,

Standard deviation = 10 mm, center height = 10 mm, height
= 40 mm

LZV510

SIP 10 sipper module set for DR 6000 complete with tray,
tube set and 1 cm quartz glass pour-through cuvette, EU

LQV157.99.30001

Advertising