Operation

23

4.6 Program

Programs are stored settings, that can be edited, stored and

retrieved. Under the PROGRAM function the following volumes are

default: 0.2 ml - 0.4 ml - 0.5 ml - 1.0 ml - 1.3 ml - 2.0 ml - 2.3 ml -

2.5 ml - 5.0 ml. When working with the mode PROGRAM, the scroll

key selects the program and it is a very fast way to switch between

different settings. Additional functions as mixing, counter etc. are also

available in the mode PROGRAM. To change the default volumes

choose EDIT.

It is also possible to replace default volumes by additional functions

(

Section 4.7 Additional functions in mode PROGRAM on page 23

).

4.7 Additional functions in mode

PROGRAM

4.7.1 MIX + PIPETTE

This function adds an automatic mixing step after normal pipetting.

1.

Select PROGRAM

EDIT, then select with the scroll key

MIX + PIPETTE and accept with OK.

2.

Then select the desired volume with the scroll key and accept

with OK.

3.

Select the pipetting speeds accordingly. After dispensing the

volume a text MIX is displayed. By pressing the trigger the

pipette starts to pipette ca. 70% of the selected volume several

times as long as the trigger is held pressed. After releasing the

trigger the pipette stops after next dispensing and BLOWOUT is

displayed.

4.

A normal blow-out function is done by pressing the trigger and

the pipette is again ready for next pipetting.

4.7.2 PIPETTE + COUNT

This function adds an automatic count number to each pipetting step.

1.

Select PROGRAM

EDIT, then select with the scroll key

PIPETTE + COUNT and accept with OK.

2.

Then select the desired volume with the scroll key and accept

with OK.

3.

Select the pipetting speeds accordingly.

4.

Next choose the maximum number of pipettings, the default

value is 999.

Note: After the maximum number of pipettings is reached, the counter
returns to zero. The counter can be reset at any time to zero by pressing
the scroll key down

OK; scroll key up

OK.