4 adjustment, 1 two point calibration – Hach-Lange TENSETTE plus User Manual Benutzerhandbuch

Seite 28

Advertising
background image

Calibration

28

Procedure:

1.

Do 10 pipettings with the minimum volume.

2.

Do 10 pipettings with the maximum volume.

3.

Calculate the accuracy (A) and precision (cv) of both series.

4.

Compare the results to the limits in the

Table 2

.

If the calculated results are within the selected limits, the adjustment

of the pipette is correct.

5.4 Adjustment

5.4.1 Two point calibration

Normally the two point calibration should be used.

1.

Do the pipetting series with both maximum and minimum

volumes.

2.

Calculate the actual results.

3.

Select MENU and select OPTIONS with the scroll key and

accept with OK.

4.

Select CALIBRATE and accept with OK.

5.

Select TWO POINTS and confirm with OK. The target maximum

and target minimum volumes are shown on the display.

6.

Select EDIT and change the actual maximum volume with the

scroll key and accept with OK.

7.

Change the actual minimum volume with the scroll key and

accept with OK.

8.

A text SAVE? is displayed.

9.

Accept with YES.

10. The adjustment has been changed.

Table 2 Maximum permissible errors according ISO 8655

Range

Volume

Accuracy

Precision

µL

µL

%

s.d. µL

cv%

0.2–5 mL

5000

±40.0

±0.8

15.0

0.3

200 ±40.0

±20.0

15.0

7.5