Section 5 calibration, 1 device requirements and test conditions, 2 calibration counter – Hach-Lange TENSETTE plus User Manual Benutzerhandbuch

Seite 27: 3 checking the calibration

Advertising
background image

27

Section 5

Calibration

All Hach TenSette plus are factory calibrated and adjusted to give the

volumes as specified with distilled or deionized water. Normally, the

pipettes do not need adjustment, but they are constructed to permit

recalibration and adjustment for liquids of different temperature and

viscosity.

5.1 Device requirements and test

conditions

An analytical balance must be used. The scale graduation value of

the balance should be chosen according to the selected test volume

of the pipette:

If the uncertainty of measurement of the balance is known, this may

be used instead of the repeatability and linearity.

Test liquid: Water, distilled or deionized, “grade 3” water conforming

ISO 3696. Tests are done in a draft-free room at a constant (±0.5 °C/

±33 °F) temperature of water, pipette and air between 15 °C (59 °F) to

30 °C (86 °F). The relative humidity must be above 50%.

5.2 Calibration Counter

By selecting MENU

OPTIONSCALIBRATECOUNTER the

number of pipetttings since last calibration is shown on the display.

The counter is reset to zero when calibration is performed.

5.3 Checking the calibration

The pipette is checked with the maximum volume (nominal volume)

and with the minimum volume. A new tip is first pre-wetted 3–5 times

and a series of ten pipettings are done with both volumes. A pipette is

always adjusted for delivery (Ex) of the selected volume. Measuring

volumes taken from balance is not allowed.

Table 1 Test Conditions

Volume range

Readable

graduation

Precision

repeatability(s)

and linearity

Uncertainty

of measure-

ment

10–100 µL

0.01 mg

0.02 mg

0.02 µL

above 100 µL

0.1 mg

0.2 mg

0.2 µL

Advertising