Section 7 troubleshooting – Hach-Lange TENSETTE plus User Manual Benutzerhandbuch

Seite 35

Advertising
background image

35

Section 7

Troubleshooting

The table below lists possible problems and their solutions.

Problem

Likely Cause

Action

Leakage

Tip incorrectly attached

Attach firmly

Foreign particles between

tip and tip cone

Clean tip cones attach

new tips

Foreign particles between

the piston, the O-ring and

the cylinder

Clean and grease O-ring

and cylinder

Insufficient amount of

grease on cylinder and

O-ring

Grease accordingly

O-ring damaged

Change the O-ring

Inaccurate

dispensing

Incorrect operation

Follow instructions

carefully

Tip incorrectly attached

Attach firmly

Calibration altered:

caused by misuse, for

example

Recalibrate according to
instructions

Inaccurate

dispensing with

certain liquids

Unsuitable calibration

High viscosity liquids may

require recalibration

Recalibrate with the

liquids in question

No dispensing

Pistons stuck or not

connected

Remove tip cone module.

Move piston by hand or

with piston removal tool.

Attach the module in

service mode.

Advertising