Operation

26

4.7.6 SEQ + ASPIRATE

The sequential aspirate mode enables serial aspirating of different

volumes.

1.

Select PROGRAM

EDIT, then select with the scroll key

SEQ + ASPIRATE and accept with OK.

2.

First choose the number of volumes (maximum 20) with the

scroll key and accept with OK. VOL 1 is displayed and the

highest possible volume is flashing.

3.

Select the first volume with the scroll key and accept with OK.

Now VOL 2 is displayed and the highest possible remaining

volume is flashing.

4.

Select the second volume with the scroll key and accept with

OK. After selecting the last volume the total volume is shown on

display and speed-in is flashing for selecting the pipettings

speeds. After selecting the speeds the first volume is shown on

the display and the pipette is ready for pipetting.

5.

Fill a clean reagent reservoir with the liquid to be dispensed.

6.

Dip the tip under the surface of the liquid and press the trigger.

This action will take up first volume and the next volume is

displayed.

7.

Withdraw the tip from the liquid touching it against the edge of

the reservoir to remove excess liquid.

8.

Dip the tip under the surface of the next liquid and press the

trigger. This action will take up the current volume and the next

volume appears on the display.

9.

Repeat steps 2. and 3. until last volume is drawn in to the tip. The

total volume is shown on the display.

10. Deliver the total volume by pressing the trigger and hold it down

to empty the tip.

11. Release the trigger to return to the ready position.
Change the tip and continue pipetting.

Note: The pipette can be emptied at any time by pressing CANCEL (left
selection key

Figure 2 Pipette operation on page 15

, item

6

)