Section 6 maintenance, 1 daily checking, 2 maintenance – Hach-Lange TENSETTE plus User Manual Benutzerhandbuch

Seite 31: 1 daily checking 6.2 maintenance, Section 6

Advertising
background image

31

Section 6

Maintenance

When the Hach TenSette plus is not in use, make sure it is stored in

an upright position. We recommend the Hach pipette stand (Cat. No.:

BBP088) for this purpose.

6.1 Daily checking

The pipette should be checked at the beginning of each day for dust

and dirt on the outside surfaces of the pipette. Particular attention

should be paid to the tip cone. No other solvents except 70% ethanol

should be used to clean the pipette.

6.2 Maintenance

If the pipette is used daily it should be checked and lubricated every

three months.

1.

The maintenance procedure starts with choosing the

Service-mode from the menu (MENU

OPTIONS

SERVICE).

2.

Press the trigger (

Figure 3 on page 32

, item

2

) before

disassembling the pipette.

3.

Pull down the complete tip cone module (

Figure 4 on page 33

).

4.

Note that the tip ejector button is in fully up position (

Figure 3 on

page 32

,

item

1

).

5.

Remove the complete tip cone module (

Figure 4 on page 33

) by

pulling it out.

6.

Seperate the tip ejector sleeves (top and bottom) by pulling it into

two parts.

7.

Push the snap fittings (

Figure 4 on page 33

,

item

6

) to release

the cylinder and remove the cylinder (

Figure 4 on page 33

,

item

4

).

8.

Push forward and clean the piston (

Figure 4 on page 33

,

item

7

)

with a dry napless cloth.

9.

Check the the complete tip cone module for foreign particles.

10. Grease the cleaned parts with the lubricant that comes with the

pipette.

11. Reassemble the the tip cone module in reverse order.

Note: In order to make the reassembling of the tip cone module easier,
position the connecting pieces of the upper and the lower part
accordingly one above the other. Press now both parts together until you
hear a “click”.

Advertising