Gb 6. installation and mounting, Installation and mounting – WIKA CPH6000 Benutzerhandbuch

Seite 28

Advertising
background image

28

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

6. Installation and mounting

6.1 Requirements for test assemblies with the CPH6000

Before starting any task, the device should be switched on briefly to

determine that there is sufficient charge in the battery. (battery level

in %). The battery level is indicated briefly after powering up in a device-

status message (see chapter 7.2.1 "Device-status message shortly

after switching on the CPH6000"). 100 % battery charge will enable

approximately 20 hours of instrument operation.

Initially the test assembly must be physically assembled and, if necessary, connected

electrically (see chapter 4.4.1 "Connection of the model CPT6000 reference pressure

sensor").

Before switching the CPH6000 on, ensure that the test assembly is not pressurised

(system is vented to atmosphere) and that the equipment is correctly assembled and in

the correct mounting position.

Only disconnect test and calibration installations once the system has

been depressurised!

Particularly small measuring ranges (e.g. < 1 bar) are orientation dependent (i.e.

the mounting position considerably influences the measurement signal). This can

be compensated, if necessary, using the TARE function (see chapter 7.4.1 "SETUP

additional menu item: Functions").

Absolute-pressure measuring ranges < 1 bar absolute are, by definition, in an overload

condition at atmospheric pressure. Therefore the message "Sensor Overflow"

appears in the display, indicating a pressure outside of the measuring range of the

attached sensor. If the pressure is reduced, so that it falls within the permissible

measuring range, then this message disappears from the display. Since absolute

pressure measuring ranges < 1 bar are always ‚overloaded‘ at atmospheric pressure,

for these measuring ranges the overload time limit function is deactivated in the SETUP

menu option: ReferenceSensor (see chapter 7.4.3 "SETUP additional menu items:

ReferenceSensor").

6. Installation and mounting

Advertising