WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

optional delay-time [sec]

Input via numeric input keypad and confir-

mation with

ENTER

With the entry of such a delay-time, the acceptance/recording of the test point is

blocked for this duration. (In the above example, after the first test point has been

recorded, 30 seconds must pass before the second test point can be recorded.)

Should all test points need to be cleared or reset, since the new

calibration consists of fewer test points than the previous test series,

the CLEAR button simply needs to be pressed. This will clear and

reset the current and all subsequent test points. (This process can take

several seconds.)

Should all recorded calibration data for all calibrations need to be

cleared at the same time see chapter 7.4.7 "SETUP additional menu

item: CLEAR CalData".