7. Commissioning, operation

a) Measuring range of the CPT6000 reference pressure sensor (which is currently

connected)

b) Current measured value of the reference pressure sensor
c) Pressure unit (settable via menu)

In MEASURING mode, at the same time as the reference-pressure value, a test

item can also be displayed on the screen (see following picture). For configuration

instructions, please see chapters 7.3.1 "MEASURING mode" and 7.3.2 "MEASURING

mode (with test item)".

Display: MEASURING with test item

a) Measuring range of the test item
b) Current measured value of the test item
c) Deviation/difference between reference and test item in the current pressure units

and in % of the measuring span (% FS) or % of measured value (% rd)

d) Pressure unit (of the test item)
e) Original output signal of the test item
f) Current value of the output signal of the test item

Pressure signal

(test item)

Electrical signal

(test item)

P

I/U

switchable