Commissioning, operation – WIKA CPH6000 Benutzerhandbuch

Seite 56

Advertising
background image

56

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

..Ref. Sensor-list

362A

R.-Start:
Range-End:
Unit:
Cal-Dat.:
Class:

0.000

10.000

bar

2011/05/10

0.025

Rp56

362A

Kd35

----

----

----

----

----

01
02
03
04
05

06
07
08
09
10

Rp52

----

Current

a

b

c

g) Ref-Value:

Current measured value of the connected CPT6000 reference pressure sensor

h) Offset:

This menu option only appears if the CPH6000’s reference pressure sensor is an

absolute pressure sensor.

Through this menu option the measured value of the reference pressure sensor

can be adjusted. This should only be used, however, as close as possible to the

absolute zero, and using a reference that is at least 4 times more accurate.

i) Cal-Dat.:

Calibration date of the CPT6000 reference pressure sensor (Year/Month/Day)

7.4.4 SETUP additional menu items: Ref. Sensor-list

7. Commissioning, operation

The ProcessCalibrator supports up to 10 CPT6000 reference pressure sensors. These

are listed in this menu.

a) Currently connected CPT6000 reference pressure sensor
b) Sensor list of the supported CPT6000 reference pressure sensors (calibrated with

the instrument)

c) Date of the sensor which has been selected using the cursor (Calibration date:

Year/Month/Day)

Selection

of a listed sensor