WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

ENTER

ENTER

7.3.5 CALIBRATION mode (calibration of a pressure transmitter)

7. Commissioning, operation

Deviation

Deviation

Deviation

No. of test item

and No. of test

point

No. of test item

and No. of test

point

No. of test item

and No. of test

point

Generate the specified

setpoint in acc. with the

reference display (establish

a pressure-free condition/

atmosphere) and with

<--

record the measured values

of this test point

Generate the specified

setpoint in acc. with the

reference display using the

pressure generator

Record the measured values

of this test point

(with

back to previous

test point)

2. Test point

x. Test point

1. Test point

(e.g. 0 bar)

2. Test point

(e.g. 1 bar)