3 installation of a wall-mounted version, Installation of a wall-mounted version, Remove blanking strips covering the screws – Burkert Type 8025 Benutzerhandbuch
Seite 18: Wire acc. to instructions in chap. “7.4, 16 installationandwiring, Figure 6: installation of a wall-mounted version, Type 8025 universal, 1106 mm, T 250 ma, 106 mm

16
Installationandwiring
7.3
Installation of a wall-mounted version
note
risk of material damage if the cable glands are not tightly screwed on the housing
▶ Before installing the wall-mounted housing on its support, tighten the nuts of the entry item of the cables
glands at a torque of 1.5 Nm.
The flow transmitter in a wall-mounted version has 4 holes in the bottom of the housing.
→
Remove blanking strips covering the screws.
Blanking strips
ENTER
0....9
FLOW
Nut of the entry item
→
Loosen the 4 screws and open the cover to get access to the holes [1].
1
1
1
1
106 mm
230V
L N
230V
T 250 mA
5 6 7 8 9 10
L N
230V
CURRENT
SOURCE
SINK
BINARY
PE
PE
PE
PE
DI1
DI2
DI3
DI4
DO4
ISOG
FLOW
SENSOR
L+
L-
PE P- P+
NC
Iout
PULSE
DO1
Supply
12..36Vdc
Univ
Batch
(AO1)
SUPPLY
NC
COIL
PULSE INPUT
NPN/PNP
2
1
3
PE
+
-
SENSOR
SUPPL
Y
LOAD
+5V
L+
(L+)-12V
COIL/PNP
39K
470
2.2K
SENSOR TYPE
COIL
NPN/PNP
DO2
DO3
OF
FO
N
106 mm
Figure 6:
Installation of a wall-mounted version
→
Secure the housing to the support respecting the dimensions indicated in “Figure 6”.
→
Wire acc. to instructions in chap. “7.4”.
→
Close the housing and tighten the 4 screws of the cover.