8.6.16 parameterizing the cut-off flow rate
This parameter makes it possible to set the flow rate value to 0 if the measured value is less than the set cut-off
value:
• the display then shows a flow rate = 0 (a dot is displayed after the flow rate units).
• the outputs and the totalizers react as if the actual flow rate was = 0.
CUt-OFF
CO=0.000
BACKLit
→
Enter the measured flow rate value below which the device considers that the flow rate
is = 0.
→
To disable the function, set CO=0.
→
Confirm.
Figure 47:
Diagram of the "CUT-OFF" parameter of the Parameters menu
→
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to save the settings or not and go back to the Process level.
8.6.17 Setting the brightness of the display and how long it stays
oN, or deactivating the backlight
This parameter makes it possible:
• to adjust the brightness of the display and how long the display is backlit after a key press.
• to deactivate the backlight.
BACKLit
BKLG= 1
END
→
Set the display brightness, from 1 to 9.
→
Confirm.
BKLG= 9
BKLG OFF
0......9
→
Or, deactivate the backlight.
→
Confirm.
tiME= 30
→
Set the duration the display is backlit after a key press (00 s: the
display is constantly backlit).
→
Confirm.
END
Figure 48:
Diagram of the "BACKLIT" parameter of the Parameters menu