Design and function, 5 tightening torques – WIKA TC12-M Benutzerhandbuch

Seite 16

Advertising
background image

GB

4. Design and function

14083458.01 09/2013 GB/D

WIKA operating instructions models TR12, TC12 (Ex i)

16

4.5 Tightening torques

Connection head, selectable (example)

Thread

Tightening torques in Nm

Connection head material

Aluminium

Stainless steel

1/2 NPT

32

35

3/4 NPT

36

40

M20 x 1.5, with counter nut

1)

23

25

M24 x 1.5, with counter nut

1)

27

30

1) only for versions with non-divisible neck tube

Tightening torques between connection head and neck tube

Tightening torques for connection to thermowell

Tightening torques for connection to neck tube

Thread

Tightening torques in Nm

R 1/2

1)

50 ... 60

1) only for versions with “nipple-union-nipple” neck tube

Thread

Tightening torques in Nm

1/2 NPT

35

3/4 NPT

40

G 1/2 B

35

G 3/4 B

40

M14 x 1.5

25 ... 30

M18 x 1.5

35

M20 x 1.5

35 ... 40

M27 x 2

40 ... 45

Only ever screw in, or unscrew, the instrument via the spanner-flats and to the prescri-

bed torque using an appropriate tool.

The correct torque depends on the dimensions of the connection thread and the gasket

used (form/material).

When screwing in or unscrewing the instrument, do not use the connection head as

contact surface.

When screwing in the instrument, please observe that the threads are not skewed.

4.6 Scope of delivery

Cross-check scope of delivery with delivery note.