Gb 3. specifications 3. specifications, Specifications – WIKA CPH6400 Benutzerhandbuch

Seite 12

Advertising
background image

12

WIKA operating instructions, model CPH6400

11313439.04 12/2012 GB/D

GB

3. Specifications

3. Specifications

Specifications

Precision hand-held pressure indicator

model CPH6400 (complete measuring chain)

Sensor technology

1 reference pressure sensor (interchangeable without tools)

1)

Measuring range

bar

0.4

1.6

6

16

40

100

250

600

1000

Overpressure safety

bar

2

10

35

80

80

200

500

1200 1500

Burst pressure

bar

2.4

12

42

96

400

800

1200 2400 3000

Accuracy of the measuring chain

0.025 % FS

2)

Measuring range

bar

1600 2500 4000 5000 6000

Overpressure safety

bar

2300 3500 5000 6000 7000

Burst pressure

bar

4000 6000 8000 10000 11000

Accuracy of the measuring chain

0.1 % FS

2)

Type of pressure

{In addition to the above-specified pressures; vacuum, bi-direction-

al ranges and absolute pressures are available}

Pressure units

bar, mbar, kPa, psi, mmHg, inHg and kg/cm² (depending on the

measuring range, selectable)

Active temperature compensation

°C

10 ... 40

Temperature measurement

only for 2-channel version

Sensor type

Pt100, 4-wire

Measuring range

°C

-10 ... +50

Resolution

°C

0.01

Accuracy

°C

0.05

Permissible ambient temperature

°C

0 ... 50

Calibration

2)

Factory calibration certificate (optional: DKD/DAkkS calibration

certificate)

Specifications

Reference pressure sensor model CPT6400

Pressure connection

≤ 1000 bar: G ½ B; {various connection adapters on request}

> 1000 bar: M16 x 1.5 female, with sealing cone

Material

Stainless steel for wetted parts

(for measuring ranges > 25 bar ... ≤ 1000 bar additional Elgiloy

®

)

Internal transmission fluid

Synthetic oil (only for measuring ranges up to 25 bar)

Permissible temperature ranges

Medium

°C

-20 ... +80

Storage

°C

-40 ... +85

Case

Stainless steel

Electrical connection

Circular connector, 8-pin

Ingress protection

IP 65 (with cable connected)

Weight

g

approx. 220

CE conformity

Pressure equipment directive

97/23/EC, PS > 200 bar; module A, pressure accessory

EMC directive

2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immuni-

ty (industrial locations)

1) Calibration data sets for up to 10 transmitters can be stored in each instrument (up to 10 calibration data sets)

2) Calibrated at 23 °C and in vertical mounting position with pressure connection facing downwards.

{} Items in curved brackets are optional extras for an additional price.

Advertising