Gb 4. design and function – WIKA CPH6400 Benutzerhandbuch

Seite 18

Advertising
background image

18

WIKA operating instructions, model CPH6400

11313439.04 12/2012 GB/D

GB

4. Design and function

When the battery is nearly empty, the message "low BAT" appears in the display. With

0 % battery level, the equipment automatically switches itself off and must then be

recharged using the battery charger.

4.5.1 During charging

CAUTION!

The temperature range over which the Lithium-Ion battery can be

charged is 10 °C ... 45 °C. Charging the Lithium-Ion battery at tempera-

tures outside of this range may lead to heating or damage. In addition,

the performance of the Lithium-Ion battery can be affected and the

service life reduced.

4.5.2 Charging the Lithium-Ion batteries

WARNING!

In order to charge the Lithium-Ion battery, never use any device other

than that specified by WIKA. When the Lithium-Ion battery is used in

devices other than the specified device, the performance and service

life of the Lithium-Ion battery may be reduced, and, should the device

cause an abnormal current to flow, it may cause the Lithium-Ion battery

to become hot, explode or ignite and result in serious injury.

CAUTION!

The temperature range over which the Lithium-Ion battery can be

discharged is -10 °C ... +60 °C. Use of the Lithium-Ion battery outside of

this temperature range may affect the performance of the battery or may

reduce its service life.

Advertising