6. Commissioning, operation

Measuring mode with MIN/MAX function

1. Battery level display

2. Current pressure measured value

3. Current unit and type of pressure

4. Minimum function

The Minimum function records and displays the lowest value since it was switched on

and is reset by switching the instrument on/off.

5. Maximum function

The Maximum function records and displays the highest value since it was switched

on and is reset by switching the instrument on/off.

6. Bar graph

MODE

Return to the standard measuring mode by pressing the MODE key several

times.

6.4.3 Measuring mode incl. the display of the pressure rate and optional tempera-

ture

Via the MODE key different functions can be superimposed, e.g. the function: Pressure

Rate (RATE).

With the 2-channel version of the CPH6400 an additional display of measured tempera-

ture can also be made, using an external Pt100 resistance thermometer. If there is no

Pt100 resistance thermometer connected with this version, then the display shows

"Waiting" instead of the temperature value.

Activation of the Pressure Rate function RATE and optional temperature display

Pressing the MODE key twice* (directly within the standard measuring mode) displays

the pressure rate and the temperature in °C.
Pressing the MODE key three times with an external Pt100 resistance thermometer

(directly within the standard measuring mode) displays the pressure rate and the temper-

ature in °F.

* If you are in a different measuring mode than the standard measuring mode, press the MODE key until the

measuring mode with the Pressure Rate function appears on the display).

1

2

3

6

5

4