Gb 6. commissioning, operation – WIKA CPH6400 Benutzerhandbuch
Seite 32
32
WIKA operating instructions, model CPH6400
11313439.04 12/2012 GB/D
GB
6. Commissioning, operation
6.8.5 Display resolution (DIGITS)
Via the FUNC key various configurations can be achieved, e.g. the display resolution can
be set.
Activating the Display Resolution menu item (DIGITS)
Pressing the FUNC key five times* (directly within the measuring mode).
* If you already are in a FUNC menu item, press the FUNC key until the "DIGITS" menu item appears.
In the "DIGITS" FUNC menu item, the displayed measured value resolution can be
decreased via the MODE key and increased via the UNIT key. In total, up to 6 digits can
be displayed.
FUNC
Return to the standard measuring mode by pressing the FUNC key several
times.
6.8.6 Zero point correction (only for gauge pressure ranges)
(OFFSET CORRECTION)
Via the FUNC key various configurations can be achieved, e.g. the zero point correction
of gauge pressure sensors.
Activation of the Zero Point Correction menu item (OFFSET CORRECTION)
Pressing the FUNC key six times* (directly within the measuring mode).
* If you already are in a FUNC menu item, press the FUNC key until the "OFFSET CORRECTION" menu
item appears.