Gb 4. design and function – WIKA CPH6400 Benutzerhandbuch

Seite 15

Advertising
background image

15

WIKA operating instructions, model CPH6400

11313439.04 12/2012 GB/D

GB

4. Design and function

4.3 Electrical connections to the CPH6400 digital indicator

All electrical connections are located along the top edge of the CPH6400 (see picture

below).

Sensor connection

Connection for interface adapter

Power connection

WARNING!

Only original WIKA components should be connected to any of the

electrical connections. (Only the WIKA battery charger to the mains

input/charger socket; and only a WIKA USB connecting cable to the

interface socket).

DANGER!

The CPH6400 must be switched off before connecting or disconnecting

any electrical connection. In addition, the supply voltage marked on the

power supply unit must match the local mains voltage.

4.4 CPT6400 reference pressure sensor

For the CPH6400, there are many reference pressure sensors to choose from, (with

measuring ranges from 400 mbar up to 1,000 bar with an accuracy of 0.025 %, and

measuring ranges over 1,000 bar up to 6,000 bar with an accuracy of 0.1 %) which can

be interchanged quickly and without tools. When the CPH6400 is switched on, the refer-

ence pressure sensor attached is recognised automatically, so that no further configura-

tion is needed.

Up to ten CPT6400 reference pressure sensors are supported by one

CPH6400.