IMPAC pyrometers series IN 5 plus

7

IN 5-H plus, IN 5/5-H plus:
0.01 s; adjustable in the pyrometer: 0.5 s; 1 s; 2 s; 5 s,

adjustable via interface: 0.5 s; 1 s; 2 s; 5 s; 10 s; 30 s

IN 5-L plus:
0.18 s; adjustable in the pyrometer: 0.5 s; 1 s; 2 s; 5 s,

adjustable via interface: 0.5 s; 1 s; 2 s; 5 s; 10 s; 30 s

Data handling:

Digital

IR-Detektor:

Thermopile

Power supply

24 V DC (18 to 30V DC) nominal, ripple must be less than 0.5 V

Power consumption:

Max. 70 mA

Analog output:

adjustable 0 to 20 mA or 4 to 20 mA (linear)

Load:

max. 500

Ω at 24 V (max. 200 Ω at 18 V)

Digital interface:

RS232 or RS485 (addressable, half-duplex)
baud rate 1200 to 19200 Bd, resolution 0.1°C

Isolation:

Power supply, analog outputs and digital interfaces are
electrically isolated from each other

Parameters:

Adjustable on the pyrometer (in offline mode):

Emissivity, exposure time, 0/4 to 20 mA analog output range,
online- / offline switch

Readable and adjustable via interface / PC (in online mode):

Emissivity, exposure time, 0/4 to 20 mA analog output range, sub
temperature range, automatic clearing of the max./min value storage,
external clearing of the max./min value storage, clear times of the
max/min value storage, address, baud rate, internal temperature °C /
°F, max. / min. selection, activation of ambient temperature correction

Emissivity:

0.2 to 1 switchable in the instrument (offline mode) or with the software

InfraWin (online mode) in steps of 0.01

Maximum value store/
Minimum value store

Built-in single and double store. Clearing with clear time t

CL

(0.1 s; 0.25 s; 0.5

s; 1 s; 5 s; 25 s), external contact or via interface or also automatically with
each new item to be measured

Sighting (optional):

Laser targeting light (max. power level < 1 mW,
λ = 630-680 nm, CDRH class II)

Weight:

410 g

Housing:

Stainless steel

Relative humidity

Non condensing conditions

Operating position:

Any

Ambient temperature:

0 to 63°C (0 to 60°C for IN 5/5 plus MB 25)

Storage temperature:

-20 to 70°C

Protection class:

IP65 (DIN 40050)

CE-label:

according to EU directives about electromagnetic immunity