WIKA Operating instructions/Betriebsanleitung D-10-9, D-11-9

ID

DLC

Byte 0

Byte 1

Byte 2

7E5h

8

15h

Byte 3

Byte 4

Byte 5

Byte 6

Byte 7

Baud Rate

Index table

1000 kBit

0

800 kBit

1

500 kBit

2

250 kBit

3

125 kBit

4

100 kBit

5

50 kBit

6

20 kBit

7

10 kBit

8

ID

DLC

Byte 0

Byte 1

Byte 2

7E5h

8

04h

00h

Byte 3

Byte 4

Byte 5

Byte 6

Byte 7

Error

Cause

Recovery action

The setting of the node-ID has been

confirmed as successful, but it is not active.

Settings are taken over

by the DIP-switch.

Change over DIP-switch so that the parameters

from the device-internal memory are loaded.

The setting of the data transfer rate has

been confirmed as successful, but it is not

active.

Settings are taken over

by the DIP-switch.

Change over DIP-switch so that the parameters

from the device-internal memory are loaded.

8. Communication (LSS-Services according to CiA 305 V1.0)

GB

Byte 1-2 Switch Delay (time in ms, which is to pass until the new Baud rate is activated).

Activate bit timing by

Bit timing table

Change from configuration mode to operation mode via switch mode global:

9. Maintenance, accessories / 10. Trouble shooting

GB

9. Maintenance, accessories

Accessories: For details about the accessories (e. g. connectors), please refer to WIKA‘s
price list, WIKA‘s product catalog on CD or or contact our sales department.

„

WIKA pressure transmitters require no maintenance.

„

Have repairs performed by the manufacturer only.

10. Trouble shooting

Open pressure connections only after the system is without pressure!

Do not insert any pointed or hard objects into the pressure port for cleaning to prevent
damage to the diaphragm of the pressure connection.

„

Take precautions with regard to remaining media in removed pressure trans-
mitters. Remaining media in the pressure port may be hazardous or

toxic!

„

Remove the pressure transmitter from service and mark it to prevent it from
being used again accidentally, if it becomes damaged or unsafe for opera-
tion.

„

Have repairs performed by the manufacturer only.

Warning

!