Initial start-up and operation gb, Operating instructions/betriebsanleitung, wud-2x – WIKA WUD-26 Benutzerhandbuch
Seite 12

Operating instructions/Betriebsanleitung, WUD-2x
11611057.01 GB/D 11/2013
12
Specifications
Models WUD-20, WUD-25 and WUD-26
Power supply U
+
DC 10 ... 30 V with output signal 4 ... 20 mA / DC 0 ... 5 V
DC 14 ... 30 V with output signal DC 0 ... 10 V
Signal output and maximum
ohmic load R
A
(in Ω)
4 … 20 mA, 3-wire, R
A
≤ (U+ – 10 V) / 0.02 A
DC 0 ... 5 V, 3-wire, R
A
> 5 k
DC 0 ... 10 V, 3-wire, R
A
> 10 k
Power P
i
1 W
Current consumption
max. 50 mA
Total current supply
max. 250 mA (incl. switching current)
Adjustability zero
-3.5 ... +3.5 % of span (via potentiometer), current output signal
-2 ... +3.5 % of span (via potentiometer), voltage output signal
Signal response (10 ... 90 %) ≤ 300 ms
Switch points
individually adjustable via external control keys
■
Type
transistor switching output NPN
■
Number
1 or 2
■
Function
normally open, normally closed, on, off
■
Switching current
Max. current that can be used with the switch
SP1: 100 mA
SP2: 100 mA
■
Accuracy
≤ 0.5 % of span
Insulation voltage
DC 500 V
Display
■
Design
7 segment-LED, red, 4-digits, height 8 mm, electronic 180° rotatable
■
Accuracy
≤ 1.0 % of span ±1 Digit
■
Update
0.2 / 0.5 / 1 / 5 / 10 / 60 sec (adjustable)
Accuracy
≤ 0.15 % of span (≤ 0.4 with pressure ranges ≤ 2 bar) RSS (Root Sum Squares)
≤ 0.3
%
1)
of span (≤ 0.6
1)
with pressure ranges ≤ 2 bar)
Non-linearity
≤ 0.1 % of span (≤ 0.15 for pressure ranges ≤ 2 bar) (BFSL) according to IEC 61298-2
Hysteresis
≤ 0.14 % of span
Non-repeatability
≤ 0.12 % of span
1-year stability
≤
0.25 % of span typ.(at reference conditions)
1) Including non-linearity, hysteresis, zero point and full-scale error (corresponds to error of measurement per IEC 61298-2)
7. Initial start-up and operation
GB