Functional test

The output signal must be proportional to the pressure. If not, this might point to a damage of the diaphragm. In

that case refer to chapter 10 „Trouble shooting“.

WARNING!

Open pressure connections only after the system is without pressure!

Observe the ambient and working conditions outlined in chapter 7 "Initial start-up and operation".

Please make sure that the pressure transducer is only used within the over load threshold limit at all times!

Menu One Key Version (without optional alarm contacts)

Operating-Mode

dISu

short press

long press to confirm setting

(seconds)

Toggle values

by short press

(relevant LED lights up)

The switch returns to the operating mode, if no key is pressed within 1 minute.

short press

long press

extra long press

(10 seconds)

Toggle units

by short press

psia MPa bar

units may vary

Rotate display

by 180°

DISU = Display Update

0.2

0.5

1

5

10

60

7. Initial start-up and operation