7. Initial start-up and operation

Mechanical connection

Product label (example)

Coded manufacture

date

Use this spanner flat for screwing in!

Don't use this spanner flat!

P# Product No.

S# Serial No.

PIN assignment

Signal

Power Supply

GB

■

Remove the protection cap only just before installation.

■

When mounting the instrument, ensure that the sealing faces of the instrument and the measuring point are

clean and undamaged.

■

Do not scratch or nick the bead. Do not over tighten. Damage to the bead will affect the fitting's performance

and may cause leakage in the system.

■

Screw in or unscrew the instrument only via the flats using a suitable tool and the prescribed torque. The

appropriate torque depends on the dimension of the pressure connection and on the sealing element used

(form/material). Do not use the case as working surface for screwing in or unscrewing the instrument.

■

When screwing the pressure transducer in, ensure that the threads are not jammed.

Mechanical Connection

■

Prepare the gas line connections appropriately.

■

You should blow clean all fitting components (such as sealing gaskets, for example) using a clean/filtered gas. Please

refer to the specific technical guides furnished by the gasket manufacturers for additional specifications.

■

You can then remove the protective film, as well as any plastic caps there may be for protecting high-quality fittings.